| Model Number |
SEE H11 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Inadequate Osseointegration (2646); No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device material number nor lot number was identified therefore complaint rate, review of device history records and risk file was not possible.The root cause for the failure cannot be determined since neither material number nor lot number were provided, and the device was not returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.Possible material number identified (b)(4).
|
| |
|
Event or Problem Description
|
|
Thoracic screws became slightly loosened from multiple bones in the same location due to unknown reasons.They were removed and replaced.
|
| |
|
Search Alerts/Recalls
|