MAUDE Adverse Event Report: TRIVIDIA HEALTH INC SYRINGES; PISTON SYRINGE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TRIVIDIA HEALTH INC SYRINGES; PISTON SYRINGE Back to Search Results
Model Number SYR, NPD 0.3CC 31G 100CT5/16" 8/CS
Medical Device Problem Codes Product Quality Problem (1506); Dull, Blunt (2407)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report reference number: (b)(4).Additional report reference numbers: (b)(4).Syringes were not returned for evaluation.Note: manufacturer contacted customer in several follow-up calls to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.
 
Event or Problem Description
Consumer reported complaint for the trueplus single-use insulin syringes; complaint was initially reported via e-mail and customer was contacted by phone.Customer reported complaint for 3 boxes of syringes: additional internal report reference numbers (b)(4).Customer stated that the unit measures printed on the syringes are "blurry" and appear to be double printed.Customer expressed concern that if the unit marking are not accurate or hard to read the insulin could be drawn incorrectly.Customer also reported they the syringe needles are not sharp and she is having a difficult time administering her medication.Customer did not have any of the syringes stating she had disposed of them.At the time of the call the customer felt well and did not report any symptoms.Customer did not claim to be injured while using the syringes and no medical intervention related to the use of the product was reported.
 
Additional Manufacturer Narrative
Sections with additional information as of 08-aug-2024: h10: syringes were not returned for evaluation.Complaint was forwarded to supplier quality based on complaint's description for investigations.No product was returned to thi.Internal evaluation has been completed by the manufacturer.No abnormalities observed with retention samples.Most likely underlying root cause onto case: mlc-009: use error caused or contributed to event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGES
Common Device Name
PISTON SYRINGE
Manufacturer (Section D)
TRIVIDIA HEALTH INC
2400 nw 55th court
fort lauderdale FL 33309
Manufacturer Contact
juliana arias
2400 nw 55th court
fort lauderdale, FL 33309
MDR Report Key19917606
Report Number1000113657-2024-00325
Device Sequence Number17953212
Product Code FMF
UDI-Device Identifier00021292005863
UDI-Public(01)00021292005863(17)280910(10)NP23131
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K993017
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 08/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberSYR, NPD 0.3CC 31G 100CT5/16" 8/CS
Device Lot NumberNP23131
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date07/12/2024
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/12/2024
Supplement Date Received by Manufacturer08/19/2024
Initial Report FDA Received Date08/06/2024
Supplement Report FDA Received Date08/28/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-