Catalog Number 395.330
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a spinal fusion surgery on (b)(6) 2024, the surgeon broke the handle on the angulates odontoid screwdriver.It was suspected that the resin-like handle was likely dried out and that is how it detached from the driver.The case was completed with the same driver.The patient was not impacted; fragments were removed easily without requiring additional intervention.The procedure was not delayed.This report involves one angled cannulated screwdriver.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: d9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: the product was not returned to depuy synthes, however photos were provided for review.The photo investigation revealed that '395.330, angled cannulated screwdriver¿ had broken handle.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the angled cannulated screwdriver would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to component failure due to unintended forces and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): part # 395.330.Synthes lot # 1010.Supplier lot # 1010.Release to warehouse date: 30 dec 1997.Supplier : (b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGLED CANNULATED SCREWDRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ  4528
Manufacturer (Section G)
ZUCHWIL : SYNTHES GMBH
luzernstrasse 19-21
zuchwil 4528
SZ   4528
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key19917730
Report Number8030965-2024-09652
Device Sequence Number8113632
Product Code HXX
UDI-Device Identifier10705034777405
UDI-Public(01)10705034777405(11)971230(10)1010
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number395.330
Device Lot Number1010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/13/2024
Initial Report FDA Received Date08/06/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/30/1997
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-