| Catalog Number |
8065772745 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Event or Problem Description
|
|
A nurse reported that during cataract surgery an ophthalmic blade was found to be dull.The surgery was completed after replacing the blade with another one.No patient impact was reported.This is the second of three reports received from the same initial reporter.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in sections h.6 and h.11.A sample was not received at the manufacturing site for evaluation for the report of dull blade with initial use; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria, therefore; the root cause for the customer complaint issue cannot be determined.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|
Search Alerts/Recalls
|