|
|
| Model Number |
CNW0T4 |
| Medical Device Problem Code |
Scratched Material (3020)
|
| Health Effect - Clinical Codes |
Iritis (1940); Swelling/ Edema (4577); Corneal Endothelial Cell Loss (4806)
|
| Date of Event |
07/08/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported during intraocular lens (iol) implant procedure, linear defect on lens was noticed.Additional information has been received, patient had suffered an iris tear, endothelial damage with swelling, prolonged surgery and issues not resolved, prognosis is good.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was returned wrapped in gauze.Solution was dried on the lens.The lens had been cut into pieces.One haptic was broken/cut gusset area, returned adhered to an optic portion.The lens pieces were cleaned with klrp to review for damage.One small optic portion near the edge was missing.A crack was observed near the center on the anterior surface on the larger returned portion.This may have been the reported linear defect.This damage may indicate a plunger override occurred.The optic also had scratches on the posterior surface near the cut areas.Other cracked areas were observed near the cut areas.It cannot be determined if scratches and cracked areas near the cut sections were caused during the removal process.The used company cartridge was also returned.Viscoelastic was observed in the cartridge.No damage was observed.The cartridge was cleaned for further evaluation.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and documentation indicated the product met release criteria.Qualified associated products were indicated.The root cause could not be determined for the reported linear defect on the lens.The lens was returned cut into pieces.A crack was observed near the center on the anterior surface on the larger returned portion.This may have been the reported linear defect.This damage may indicate a plunger override occurred.The optic also had scratches on the posterior surface near the cut areas.Other cracked areas were observed near the cut areas.It cannot be determined if scratches and cracked areas near the cut sections were caused during the removal process.The used company cartridge was also evaluated.No cartridge damage was observed.Top coat dye stain testing was conducted with acceptable results.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The handpiece ifu instructs: important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company.Verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately ½ turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The ifu instructs: follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|