The q-stress device is intended to acquire, process, record, archive, analyze, and output electrocardiographic data during physiologic stress testing.The device is intended for use in adult, adolescent, and children patient populations.The device is intended for use in a clinical setting by trained personnel under the supervision of a licensed physician.The device may interface with equipment including a treadmill or ergometer for dynamic exercise evaluation, as well as non-invasive blood pressure equipment, functional arterial oxygen saturation (spo2) equipment, and computer communications equipment.Log files were reviewed and it was determined that the reported problem was due a software issue.Technical services walked the customer through how to reconcile the exam data to the correct patient.Based on this information, no further actions are required at this time.Although there was no injury reported as a result of this event, if patient data were to be mis matched on the system or an exam, it could lead to serious injury or death.Therefore, hillrom is reporting this malfunction.
|