| Model Number |
9847 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Other, other text: the returned device was investigated and it was confirmed there was a short inside the power button.This resulted in the on/off button being activated when not physically pressed.Masimo initiated a recall for this issue and notified us fda on 02/15/2024.The recall number is z-1537-2024.
|
| |
|
Event or Problem Description
|
|
The customer reported that the rad-g does not stay on continuously.There was no patient impact or consequence reported.
|
| |
|