Model Number 9847
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Other, other text: the returned device was investigated and it was confirmed there was a short inside the power button.This resulted in the on/off button being activated when not physically pressed.Masimo initiated a recall for this issue and notified us fda on 02/15/2024.The recall number is z-1537-2024.
 
Event or Problem Description
The customer reported that the rad-g does not stay on continuously.There was no patient impact or consequence reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-G
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
tahereh sedighi
15750 alton pkwy
irvine, CA 92618
9492977862
MDR Report Key19922341
Report Number3019388613-2024-00106
Device Sequence Number8113807
Product Code DQA
UDI-Device Identifier00843997013284
UDI-Public(01)00843997013284(11)220715
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K201770
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 07/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9847
Device Catalogue Number9847
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/22/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2024
Initial Report FDA Received Date08/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/15/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1537-2024
Patient Sequence Number1
Patient SexUnknown
-
-