| Model Number |
EPIQ 7 DIAGNOSTIC ULTRASOUND SYSTEM |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Sending supplemental report to populate become aware date in the 'date received by manufacturer' field for the initial report received by fda on july 09, 2024.This date was inadvertently left blank.
|
| |
|
Event or Problem Description
|
|
A customer reported the swivel mechanism of their epiq 7cv ultrasound system¿s control panel did not lock properly while transporting the unit within the user facility.The failure occurred outside of clinical use, and no patient or user was harmed as a result of the issue.The philips service engineer repaired the system by reseating and securing the bushing.A thorough investigation was performed to identify the root cause of the reported issue.The engineering team determined the failure was caused by a hardware issue in the articulating arm's latching mechanism, which prevented the locking solenoid from fully engaging.
|
| |
|