The device was not returned for analysis.Electronic lot history record (elhr) and corrective and preventive actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint handling database identified no other incidents from this lot.The reported patient effect of myocardial infarction is listed in the xience skypoint everolimus eluting coronary stent system instructions for use as a known adverse event which may be associated with pci treatment procedures and the use of a coronary stent use in native coronary arteries.A conclusive cause for the reported myocardial infarction, and the relationship to the product, if any, cannot be determined.The treatment appears to be related to the operational context of the procedure.Based on the reported information and results of the complaint investigation there is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
It was reported that on (b)(6) 2024, the patient, with a history of coronary artery disease and non st elevation myocardial infarction (nstemi) had one xience skypoint (3.25x28 mm) stent implanted.On (b)(6) 2024, the patient was re-admitted with chronic ischemic heart disease and stemi.Balloon angioplasty / thrombectomy was performed to treat the condition.There was no adverse patient sequela.The patient was discharged home on 05/03/2024.No additional information was provided.
|