Internal reference number: (b)(4).H10: smith & nephew is submitting this report pursuant to the provisions of 21 c.F.R.Part 803.This report may be based upon information which smith & nephew has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, smith & nephew, or its employees, that the report constitutes an admission that the device, smith & nephew or its employees caused or contributed to the potential event described in this report.
|
|
It was reported that, after a patient was implanted a bhr resurfacing femoral head 48mm, they had a metal allergy and metallosis complications.A revision surgery is currently being rescheduled.Patient's current health status is unknown.Further complications were not reported.
|
|
H3, h6: it was reported that, after a patient was implanted a bhr resurfacing femoral head 48mm, they had a metal allergy and metallosis complications.A revision surgery is currently being rescheduled.Patient's current health status is unknown.Further complications were not reported.As of today, the devices, all of which were used in treatment, remain implanted and therefore cannot be tested.Without a definitive batch number, a complete review of the historical complaints data cannot be performed for the alleged devices.A review of historical complaints data was performed using the part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the head, and this failure will continue to be monitored.As no device batch numbers were provided for investigation, manufacturing record review could not be performed.If more information is received, this investigation will be reopened.The review of the most recent ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.No further escalation actions are required.To the date, no clinically sufficient data was provided to perform a medical investigation.Based on the information provided, further investigation of the reported complaint cannot be carried out and remains inconclusive.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
|