| Catalog Number |
62422603 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a automatrix shaft broke during use.No injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned product was 1flexshaft date code 1223 with coil deformation/weld failure causing the product to not function properly.Capa-2023-519 opened to address automatrix flexshaft assemblies breaking for product manufactured by sarasota since september 2022 (date code 0922) through implementation date of may 2024 (date code 0524).Complaint is considered substantiated.Dhr nor retain evaluation can be conducted as there is no batch information provided in case.
|
| |
|