• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC AUTOMATRIX SHAFT; INSTRUMENTS, DENTAL HAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC AUTOMATRIX SHAFT; INSTRUMENTS, DENTAL HAND Back to Search Results
Catalog Number 62422603
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received where this malfunction resulted in a serious injury.Therefore, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event or Problem Description
In this event it is reported that a automatrix shaft broke during use.No injury occurred.
 
Additional Manufacturer Narrative
Returned product was 1flexshaft date code 1223 with coil deformation/weld failure causing the product to not function properly.Capa-2023-519 opened to address automatrix flexshaft assemblies breaking for product manufactured by sarasota since september 2022 (date code 0922) through implementation date of may 2024 (date code 0524).Complaint is considered substantiated.Dhr nor retain evaluation can be conducted as there is no batch information provided in case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOMATRIX SHAFT
Common Device Name
INSTRUMENTS, DENTAL HAND
Manufacturer (Section D)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer (Section G)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer Contact
dan eagar
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key19932711
Report Number2515379-2024-00039
Device Sequence Number16828255
Product Code DZN
UDI-Device IdentifierD010624226011
UDI-PublicD010624226011
Combination Product (Y/N)N
Initial Reporter CountryUK
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Consumer
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 10/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number62422603
Device Lot Number1223
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date08/06/2024
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/06/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/07/2024
Supplement Report FDA Received Date10/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-