| Catalog Number |
3505-6545 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that four sequoia closure tops stripped intra-operatively, and it was found that the tulip head of the bone screw disassembled from the shaft.The procedure was completed by replacing the implants with another closure top and screw.This resulted in a 30 minute procedural delay, but there was no patient impact.This is report five of five for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2024-00208 through 3012447612-2024-00212.
|
| |
|
Event or Problem Description
|
|
It was reported that four sequoia closure tops stripped intra-operatively, and it was found that the tulip head of the bone screw disassembled from the shaft.The procedure was completed by replacing the implants with another closure top and screw.This resulted in a 30 minute procedural delay, but there was no patient impact.This is report five of five for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tulip has disassembled from the screw shank.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to off-axis forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
|
| |
|