MAUDE Adverse Event Report: ZIMMER BIOMET SPINE, INC. CANNULATED 5.5 POLYAXIAL SCREW 6.5X45; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE, INC. CANNULATED 5.5 POLYAXIAL SCREW 6.5X45; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3505-6545
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that four sequoia closure tops stripped intra-operatively, and it was found that the tulip head of the bone screw disassembled from the shaft.The procedure was completed by replacing the implants with another closure top and screw.This resulted in a 30 minute procedural delay, but there was no patient impact.This is report five of five for this event.
 
Additional Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference reports 3012447612-2024-00208 through 3012447612-2024-00212.
 
Event or Problem Description
It was reported that four sequoia closure tops stripped intra-operatively, and it was found that the tulip head of the bone screw disassembled from the shaft.The procedure was completed by replacing the implants with another closure top and screw.This resulted in a 30 minute procedural delay, but there was no patient impact.This is report five of five for this event.
 
Additional Manufacturer Narrative
Device evaluation: the returned device matches the information in the complaint file and was examined.Visual inspection revealed the tulip has disassembled from the screw shank.Root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to off-axis forces applied during use.Dhr review: the dhr was reviewed.There are no indications of manufacturing issues which would have contributed to this event and the device was likely conforming when it left highridge medical¿s control.Device usage: this device is used for treatment.If additional information is obtained that adds value to the relevant content of this report, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULATED 5.5 POLYAXIAL SCREW 6.5X45
Common Device Name
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE, INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer Contact
teresa george
10225 westmoor dr.
westminster, CO 80021
3035018506
MDR Report Key19935086
Report Number3012447612-2024-00212
Device Sequence Number4876273
Product Code NKB
UDI-Device Identifier00889024335707
UDI-Public(01)00889024335707(10)88YT
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K132884
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3505-6545
Device Lot Number88YT
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/07/2024
Supplement Report FDA Received Date01/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/18/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexPrefer Not To Disclose
Patient EthnicityNon Hispanic
-
-