| Model Number |
MMT-441AH |
| Medical Device Problem Code |
Loss of or Failure to Bond (1068)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial and final mdr (b)(4) - device 2 of 2 e1: patient city: (b)(6).Patient country: canada.
|
| |
|
Event or Problem Description
|
|
Reference number (b)(4).Event occurred in canada.It was reported on 06-jul-2024 two infusion set tape was not sticking and set has been used for 3 days.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary: "trend observed: increase in complaints related to adhesive on ewis" has been opened on 18-sep-2024 to address all adhesive issues related to ewis product family as no specifications exist for the adhesive used on this product (design related capa).
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|
Additional Manufacturer Narrative
|
|
Supplemental report 02 - (b)(4).Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary.Capa-2010953 "trend observed: increase in complaints related to adhesive on ewis" has been opened on 18-sep-2024 to address all adhesive issues related to ewis product family as no specifications exist for the adhesive used on this product (design related capa).
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|