Model Number MPI
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 11/15/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Failure to osseointegrate.
 
Additional Manufacturer Narrative
Circumstances are not clear due to the lack of provided information.To date, no information has indicated that the device or the way it was used has contributed to the described event.In addition, no signals were found indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to osseointegration problem, and escaping the available detections has been assessed and concluded to be unlikely in high level of confidence.If additional information is received indicating otherwise, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOLECULAR PRECISION IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
DITRON DENTAL LTD
2 haofe st.
southern industrial zone
ashkelon, israel 78780 02
IS  7878002
Manufacturer (Section G)
DITRON DENTAL LTD
2 haofe st.
southern industrial zone
ashkelon, israel 78780 02
IS   7878002
Manufacturer Contact
2 haofe st.
southern industrial zone
ashkelon, israel 78780-02
MDR Report Key19936648
Report Number3010948854-2021-00161
Device Sequence Number15307028
Product Code DZE
UDI-Device Identifier07290109280349
UDI-Public07290109280349
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K140728
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/08/2026
Device Model NumberMPI
Device Catalogue Number13420
Device Lot NumberMPI001000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/02/2021
Initial Report FDA Received Date08/08/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-