| Model Number |
GLENOID, PE WITH PEG, MEDIUM |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
08/22/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that after an initial eclipse surgery in (b)(6) 2012 a revision surgery of the medium pegged glenoid was necessary after may 2017 due to a glenoid loosening.No further information received.Update dw 06-aug-2024: further information was provided that the revision surgery of the ar-9105-02 with the lot 1147008, which was implanted on (b)(6) 2012, was performed on (b)(6) 2017.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to a patient-specific event.
|
| |
|