| Model Number |
A22040T |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the sheath had a broken ceramic tip.There were no reports of patient harm.
|
| |
|
Event or Problem Description
|
|
The issue was foun during maintenance.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the device evaluation and legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 3 years since the subject device was manufactured.The device was returned to olympus for inspection, and the malfunction was confirmed.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the damage was thermally and mechanically induced.Olympus will continue to monitor field performance for this device.
|
| |
|