The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.Based on the information available, the device was intentionally used outside the way it was designed and intended.The performance, safety, efficacy, longevity, and durability of rings and bioprosthetic valves have not been determined when the product is used outside the scope of its intended use and/or its studied population.Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction related to a manufacturing deficiency.The instructions for use (ifu) have been reviewed and no inadequacies have been identified with regard to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The most likely cause is patient factors, including implant position and patient's age at time of implant.
|