MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS LEGEND PNEUMATIC MOTOR; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS LEGEND PNEUMATIC MOTOR; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number PM200
Medical Device Problem Codes Detachment of Device or Device Component (2907); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2024
Type of Reportable Event Malfunction
Event or Problem Description
Repair request initiated for device with the report of noise.It was reported there was no patient involvement.Repair request escalated to a product event due to evaluation finding of detached collet.
 
Additional Manufacturer Narrative
H3:product analysis (b)(4): evaluation determined collet detachment.The likely cause of failure is bearing worn.I was also noted hose cut, motor worn, collet broken/cracked.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND PNEUMATIC MOTOR
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key19940577
Report Number1625507-2024-00264
Device Sequence Number4877499
Product Code HBB
UDI-Device Identifier00763000430979
UDI-Public00763000430979
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPM200
Device Catalogue NumberPM200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2024
Initial Report FDA Received Date08/08/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/08/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-