| Catalog Number |
0279351400 |
| Medical Device Problem Code |
Continuous Firing (1123)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the device continuously activated.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the device continuously activated.
|
| |
|
Additional Manufacturer Narrative
|
|
Alleged failure: 90-s rf probe actived by itselt without triggered, the failure(s) identified in the investigation is consistent with the complaint record.The probable root cause/s could be a probe short due to a fluid ingress, an internal leak due to a broken/damaged polyimide bond, and suction tubing.Tissue residue inside the electrode suction path hole.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.
|
| |
|