MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS ATT MR8 ANGLED; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS ATT MR8 ANGLED; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number MR8-AA14-R
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Codes Burn(s) (1757); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: no conclusion can be drawn.Evaluation could not be performed as the device was not returned.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that druing the surgery the doctor's hand had burns due to device overheating.Additional information regarding the patient impact was being followed up from the facility.
 
Event or Problem Description
On follow-up it was reported that no injuries during the case and no patient impact and no one was burned by the drill.
 
Additional Manufacturer Narrative
H3: product analysis: evaluation could not be performed because of bearing detached.It was also noted spacer corroded.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATT MR8 ANGLED
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key19947789
Report Number1625507-2024-00268
Device Sequence Number7207866
Product Code HBB
UDI-Device Identifier00763000063443
UDI-Public00763000063443
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
PMA/510(K) Number
K183515
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AA14-R
Device Catalogue NumberMR8-AA14-R
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2024
Supplement Date Received by Manufacturer08/13/2024
Initial Report FDA Received Date08/09/2024
Supplement Report FDA Received Date09/10/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/10/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-