The device was returned to olympus.And the evaluation found loss of watertightness of the camera cable.Based on the results of the investigation, it is likely, that the following led to the malfunction: the most probable cause of the complaint was, due to damage on the cable watertightness is lost.Since, the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|