| Catalog Number |
121928050 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Pain (1994); Discomfort (2330)
|
| Date of Event |
07/30/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).D4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
A patient had revision total hip replacement surgery in (b)(6) 2024 as a result of a dislocation and associated liner dissociation.The components implanted at the time of revision were: 28/50 neutral marathon liner (121928050) and a 28mm +1.5mm articuleze metal femoral head (136511000).On saturday, july 27th, patient was getting up from lying on the sand at the beach and felt something change in their hip.Patient then experienced pain and went to the hospital.After x-rays were taken it was determined revision surgery needed to take place.Patient had surgery and it was determined that the liner (121928050) had dissociated from the acetabular shell (121732050).Surgeon removed the dissociated liner (121928050) and removed the femoral head (136511000).Hip was reduced and closure took place.All went according to plan and finished successfully.Explanted components were retain and sent to local implant retrieval lab for analysis, as per regional protocols.No additional information is available.
|
| |
|
Event or Problem Description
|
|
Additional information was received: there was no surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added : b5, d4 lot corrected: d4 primary udi number.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: an analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.
|
| |
|