| Catalog Number |
826828 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
|
| |
|
Event or Problem Description
|
|
11 of 12 reports (different patients, same failure).Other mfg report numbers: 3013886523-2024-00228, 3013886523-2024-00229, 3013886523-2024-00230, 3013886523-2024-00231, 3013886523-2024-00232, 3013886523-2024-00233, 3013886523-2024-00234, 3013886523-2024-00235, 3013886523-2024-00236, 3013886523-2024-00237, 3013886523-2024-00239.Case 6: a facility reported a directlink module (id 826828) with inconsistent values.The module was used with cerelink sensor (id 826851).The sensor was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Directlink module was returned for evaluation: dhr - lot 179944 sn (b)(6), conformed to the specifications when released to stock.Failure analysis - the module was cleaned with isopropyl alcohol 70% and visually inspected: no defect was noted.The module was inspected according to procedures: no defect was noted.Root cause - no exact root cause could be determined as the technician could not confirm any problem with the device at the time of investigation.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected field: g3.Updated us regulatory awareness date to (b)(6) 2024.Previous follow-up report was sent with incorrect date: (b)(6) 2024.
|
| |
|