1 of 12 reports (different patients, same failure).Other mfg report numbers: 3013886523-2024-00229, 3013886523-2024-00230, 3013886523-2024-00231, 3013886523-2024-00232, 3013886523-2024-00233, 3013886523-2024-00234, 3013886523-2024-00235, 3013886523-2024-00236, 3013886523-2024-00237, 3013886523-2024-00238, 3013886523-2024-00239.Case 1: a facility reported a directlink module (id 826828) with inconsistent values.The module was used with cerelink sensor (id 826851).The sensor was explanted.
|
|
The directlink icp module was returned for evaluation: dhr - lot 179944, sn (b)(6), conformed to the specifications when released to stock.Failure analysis - the module was cleaned with isopropyl alcohol 70% and visually inspected: no defect was noted.The module was inspected according to procedures: no defect was noted.Root cause - no exact root cause could be determined as the technician could not confirm any problem with the device at the time of investigation.
|