MAUDE Adverse Event Report: WESTMED LLC PULSET; PULSET 3CC SYRINGE 22G X 1" (.7 X 25MM) Q-CORK, 25U BH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WESTMED LLC PULSET; PULSET 3CC SYRINGE 22G X 1" (.7 X 25MM) Q-CORK, 25U BH Back to Search Results
Model Number 3302
Medical Device Problem Codes Fluid/Blood Leak (1250); Infusion or Flow Problem (2964)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
All information reasonably known as of 09 aug 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an airlife product is defective or caused serious injury.
 
Event or Problem Description
Some of the blood overflows from the syringe. .
 
Additional Manufacturer Narrative
All information reasonably known as of 09 aug 2024 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by airlife.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to airlife.Airlife.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the airlife complaint database and identified as complaint 10101.This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an airlife product is defective or caused serious injury.The complaint of "some of the blood overflows from the syringe" regarding part 3302+ was not confirmed.The root cause could possibly be a result of a molding failure with flash or that the syringe is dimensionally inconsistent.A risk assessment was performed, and the ultimate risk was determined to be medium which does require reporting to the capa review board.There has been 1 other complaint regarding the same part and a similar issue within the 24 months preceding the reporting of this issue.A resolution letter was sent to the customer.Complaints will continue to be monitored for possible trends.
 
Event or Problem Description
Some of the blood overflows from the syringe. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSET
Common Device Name
PULSET 3CC SYRINGE 22G X 1" (.7 X 25MM) Q-CORK, 25U BH
Manufacturer (Section D)
WESTMED LLC
5580 s nogales highway
tuscon AZ 85706
Manufacturer (Section G)
WESTMED JAVID DE MEXICO, S.A. DE C.V
carretera internacional km 6.5
edif 18; parque industrial nog
nogales sonora, mexico 84094
MX   84094
Manufacturer Contact
melissa brickley
2710 northridge dr nw suite a
grand rapids, MI 49544
6162598415
MDR Report Key19955343
Report Number2028807-2024-00018
Device Sequence Number13002412
Product Code CBT
UDI-Device Identifier00709078011887
UDI-Public(01)00709078011887(10)447799
Combination Product (Y/N)N
Initial Reporter CountryCH
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date (Section B) 08/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3302
Device Catalogue NumberN/A
Device Lot Number447799
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/09/2024
Supplement Report FDA Received Date08/22/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-