|
Manufacturer's investigation conclusion: the reported event of the heartmate system monitor (serial number: (b)(6) not responding to touch input could not be confirmed through this analysis.Provided information indicated that the system monitor was rebooted before use.Although the system monitor was not returned for evaluation, a corrective and preventative action (capa) was opened to further investigate issues related to the system monitor atypical screen behavior.The device history records were reviewed, and the records revealed the system monitor (serial number: (b)(6) was manufactured in accordance with manufacturing and qa specifications.Heartmate 3 patient handbook section 5, entitled ¿alarms and troubleshooting¿, and heartmate 3 instructions for use section 7, entitled ¿alarms and troubleshooting¿, cover all alarms (visual and audible) and the actions to take if the alarms cannot be resolved.The heartmate 3 ifu section 4, entitled ¿system monitor¿, instructs the user on how to setup and use the system monitor with the heartmate power module.Heartmate 3 patient handbook section 6, entitled "caring for the equipment", describes how to care for and clean all equipment, including the system monitor.Section 10, entitled ¿safety checklists¿, provides checklists to assist the patient in performing routine maintenance of heartmate 3 lvad.Heartmate 3 patient handbook cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.The device is included in the system monitor atypical screen behavior advisory issued by abbott on 08 may 2024.No further information was provided.The manufacturer is closing the file on this event.
|