| Model Number |
SPL-FS |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the ultrasonic lithotriptor experienced intermittent activation.The issue occurred during an unspecified procedure.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the legal manufacturer's final investigation.The device was returned to olympus for inspection and the reported failure was not confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device malfunction was not confirmed during evaluation, the definitive root cause of the reported issue could not be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was reported the ultrasonic lithotriptor experienced intermittent activation.The issue occurred during an unspecified procedure.There were no reports of patient harm.
|
| |
|