Catalog Number 02-022-44-3515
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Emotional Changes (1831); Pain (1994); Synovitis (2094)
Date of Event 04/05/2024
Type of Reportable Event Serious Injury
Event or Problem Description
Legal case ¿ usa.It was reported that approximately 58 months after a right total knee replacement procedure, the patient underwent a revision procedure to address prosthesis wear, polyethylene wear, encapsulated polyethylene debris, synovectomy.Revision surgery operative notes were provided.Patient left the operating room in stable condition.Post operative diagnosis noted wear of the implant.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated: h6: type of investigation the following sections were corrected: h6: health effect - clinical code the reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB IMP PSC INSERT SZ 3.5, 15MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC.
2320 nw 66th ct
gainesville FL 32606 6841
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key19959952
Report Number1038671-2024-02771
Device Sequence Number20603126
Product Code JWH
UDI-Device Identifier10885862305237
UDI-Public10885862305237
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K152170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number02-022-44-3515
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/06/2024
Supplement Date Received by Manufacturer11/19/2024
Initial Report FDA Received Date08/12/2024
Supplement Report FDA Received Date11/22/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/04/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexUnknown
-
-