| Model Number |
DRILL BIT, 1.7MM |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/10/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
On 16 july 2024, it was reported by a sales representative via email that both an ar-8916-14 drill bit, 1.7mm and an ar-18700-54 driver shaft, t8, self retaining, ao, broke during use.This occurred on (b)(6) 2024 in flinders medical centre during a navicular orif.Using arthrex equipment for open reduction internal fixation 1.7mm drill bit (ref: ar-8916-14) snapped at drill tip.2.4mm t8 screwdriver (ref: ar-18700-54) snapped while screwing locking screw into locking plate.Ar-18724p-54 plate removed from patient as screw thread secured into plate and unable to be removed.All arthrex stock removed and discarded.The case was completed successfully using a synthes va foot set.There was case involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to improper bone prep; misaligned insertion; prying/leveraging the device during use.
|
| |
|