Model Number AA15
Medical Device Problem Codes Overheating of Device (1437); Loss of Power (1475)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Repair request initiated for device with the report of does not work and overheating.No patient impact reported.Repair is being escalated to product event due to reason for the return.
 
Additional Manufacturer Narrative
H3: product analysis((b)(4)): device evaluation anticipated, but not yet begun.Date of notification: 16-july-2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3: an overheating test could not be conducted due to another device malfunction.The likely cause was identified as it's not possible to lock a burr.It was also noted bearings detached, unlock/off lock ring does not turn, color ring is scratched, lock cover sticking.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND ATTACHMENT
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key19961717
Report Number1625507-2024-00280
Device Sequence Number7136547
Product Code HBB
UDI-Device Identifier00763000084653
UDI-Public00763000084653
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAA15
Device Catalogue NumberAA15
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/12/2024
Supplement Report FDA Received Date10/14/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-