| Model Number |
20000000100011360 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during functional testing, the device prompted a "unit failed" message.Complainant did not indicate that there was any patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Justification for no udi, this device was manufactured but not domestically distributed; it is only distributed in the intended geography.There is no existing udi regulation.Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was observed during initial testing; however, after disassembling the device to determine root cause, the report was no longer seen.The high voltage capacitors will be replaced as a precaution.The device will be recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|