• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION; GENERATOR, LESION, RADIOFREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION; GENERATOR, LESION, RADIOFREQUENCY Back to Search Results
Model Number CSK-TC10
Medical Device Problem Code Structural Problem (2506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/17/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.
 
Event or Problem Description
It was reported that the patient underwent a radiofrequency ablation (rfa) procedure during which an electrode became detached from the connector while inside the patient.The outcome of the procedure was unable to be obtained.
 
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block 26b: gxd, gxi.
 
Event or Problem Description
It was reported that the patient underwent a radiofrequency ablation (rfa) procedure during which an electrode became detached from the connector while inside the patient.The outcome of the procedure was unable to be obtained.
 
Event or Problem Description
It was reported that the patient underwent a radiofrequency ablation (rfa) procedure during which an electrode became detached from the connector while inside the patient.The outcome of the procedure was unable to be obtained.After further review, the broken electrode was discovered post-procedure and did not occur inside the patient.As such, there was no reported patient harm.
 
Additional Manufacturer Narrative
Device analysis performed on the returned electrode revealed normal device characteristics during visual and functional testing.Therefore, with all the available information, engineers concluded that the cause of the reported event was unable to be determined.B3: approximated based on the date the manufacturer became aware of the event.26b: gxd, gxi.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
GENERATOR, LESION, RADIOFREQUENCY
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key19967036
Report Number3006630150-2024-05264
Device Sequence Number18367067
Product Code GXD
UDI-Device Identifier00813250010107
UDI-Public00813250010107
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K050084
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCSK-TC10
Device Catalogue NumberCSK-TC10
Device Lot Number31746134
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2024
Supplement Date Received by Manufacturer10/17/2024
12/06/2024
Initial Report FDA Received Date08/12/2024
Supplement Report FDA Received Date11/11/2024
12/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/02/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;