| Catalog Number |
3910-500-422 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the needle separated during the procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that the needle separated during the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: misfire.Probable root cause: process: improper swaging of needle onto suture.Application: excessive force.The reported failure mode will be monitored for future reoccurrence.
|
| |
|
Search Alerts/Recalls
|