• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER INTERNATIONAL INC. INTRAVIA; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER INTERNATIONAL INC. INTRAVIA; CONTAINER, I.V. Back to Search Results
Model Number 2J8002
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/11/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
E1: initial reporter address - (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an unspecified quantity of intravia 250 ml capacity containers had large air bubbles appear in the bags.The bags are aired out completely to prevent ¿air-in-line¿ pump alarms when used by the patient; however, when checking the bags, small air bubbles start to appear from the drug additive port and coalesce into big air bubbles.Patients would have to re-prime the line to remove the air bubbles.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H11: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The devices were not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRAVIA
Common Device Name
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER INTERNATIONAL INC.
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal 91000
DR   91000
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key19968565
Report Number1416980-2024-04218
Device Sequence Number8962037
Product Code KPE
UDI-Device Identifier00085412029573
UDI-Public(01)00085412029573(10)DR24B20055
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K964853
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2J8002
Device Catalogue Number2J8002
Device Lot NumberDR24B20055
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2024
Supplement Date Received by Manufacturer09/10/2024
Initial Report FDA Received Date08/12/2024
Supplement Report FDA Received Date09/11/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INOTROPE
Patient SexUnknown
-
-