|
|
| Catalog Number |
EP8747H |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H6 component code: g07002 pending device evaluation the product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During the procedure, the suture broke when trying to tie the knot.A new box with a different lot number was used to complete the case without patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional h6 component code: g07002 no device problem found.Investigation summary of device from same lot/batch returned by user: the product was returned to ethicon for evaluation.One sealed box with thirty-six representative unopened samples was received for analysis.Product code ep8747h.The complaint sample was not received for evaluation.Upon initial inspection of the samples, no external damages were observed on the external packets.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issue related to the customer complaint and no defects were observed during evaluation.A functional test was performed using instron equipment and tensile force was above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|