| Model Number |
620-0221510-01 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer was contacted for return of the suspect product.The product has not been returned to zoll for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation for evaluation.The customer's report was duplicated and attributed to a faulty transient voltage suppressor on the dock connector board.The dock connector board will be replaced to resolve the report.The device will be recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|