• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Catalog Number GSX0020A
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Migration (4003)
Health Effect - Clinical Code Foreign Body Embolism (4439)
Date of Event 07/24/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported the physician implanted a 20mm gore® cardioform septal occluder to close a multi fenestrated atrial septal defect.The physician was attempting to deploy this smaller device in the smaller of the two defects.Unknowingly the device was advanced through the larger defect.The device was locked and released.Shortly after removing the retrieval cord the device embolized.The device was removed with a snare from the abdominal aorta.The physician stopped the case and will reassess the patient at a later date.The patient was doing well following the procedure.
 
Additional Manufacturer Narrative
Per the gore® cardioform asd occluder instruction for use, in the section "potential device ¿ or procedure-related adverse events".Adverse events associated with the use of the occluder may include but are not limited to: device embolism.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may have not been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
Received user facility medwatch.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
dan kitterman
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key19972462
Report Number2017233-2024-05212
Device Sequence Number15550289
Product Code MLV
UDI-Device Identifier00733132631018
UDI-Public00733132631018
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
P050006
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/07/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/30/2024
Device Catalogue NumberGSX0020A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/13/2024
Supplement Report FDA Received Date11/07/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age48 YR
Patient SexMale
-
-