| Model Number |
T10 HEXALOBE, CMP FT |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/31/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during a foot surgery the tip broke off inside the patient.The broken pieces were retrieved from patient.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.Complaint allegation was confirmed.One packaged ar-8737-38 batch 1392315 was received for evaluation.Visual inspection noted that the tip of the instrument was broken off.No functional testing was performed as the instrument was broken off.The instrument's overall length is shorter due to the breakage.Drawing c4943 at revision 13 measurement: 5.0 ±.03 result: 4.84 measurements at the lower place of the breakage.The most likely cause of the reported failure is a user error, as mechanical damage to the device such as hitting the device with another device or object, prying/leveraging, or excessive bending forces applied during use.
|
| |
|