| Model Number |
CV-V1 |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The evaluation of the event is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported, that the video system center had exhibited an image issue, which rendered the endoscopic image non-visible.The issue was observed during maintenance.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 10 years since the subject device was manufactured.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the reported issue could not be determined.Olympus will continue to monitor field performance for this device.
|
| |
|