|
MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3502650 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
07/22/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient who underwent breast augmentation revision with a mentor smooth round moderate plus profile 650cc saline prosthesis experienced left sided deflation post procedure.As a result, explant is planned for (b)(6) 2024.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Future explant date: (b)(6) 2024.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was received on september 26, 2024.Implant replacement was performed with explant.The impacted product was received by the failure analysis lab on october 14, 2024.The investigation of the returned device was completed by the failure analysis lab on october 15, 2024.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the breast implant had a crease/fold on the posterior view.In addition, a tear was noticed within the crease/fold measuring approximately 0.9 cm.The evaluation determined that the possible cause of the rupture was consistent with normal wear.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in deflation at a later time.Breast implants may also simply wear out over time.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|