| Model Number |
DBEC-125 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Myocardial Infarction (1969)
|
| Date of Event |
03/12/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 investigation conclusion code 22: the diamondback 360® coronary orbital atherectomy system instructions for use manual states that myocardial infarction is a potential adverse event that may occur and/or require intervention with use of the system.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The device history record for the reported oad could not be reviewed as the lot number was not provided.Csi id: (b)(4).
|
| |
|
Event or Problem Description
|
|
Following atherectomy using diamondback 360 coronary orbital atherectomy device (oad), angiographic imaging observed minor dissection in mid left anterior descending artery.Additional information was received from the clinical event committee (cec).The cec reviewed procedural images and concluded that the diamondback 360 coronary orbital atherectomy device possibly contributed to myocardial infarction.
|
| |
|