EXACTECH, INC. LOGIC TIB INSERT IMPL CRC, SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Catalog Number |
02-012-51-4011 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Emotional Changes (1831); Pain (1994); Loss of Range of Motion (2032); Swelling/ Edema (4577)
|
| Date of Event |
01/31/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant products: (b)(6) - 02-010-03-0240 - logic cr femoral cem, left, sz 4.(b)(6) - 02-012-45-4040 - lgc tibial fit tray cem sz 4f / 4t.(b)(6) - 200-02-38 - three peg patella 38mm.(b)(6) - 201-78-82 - collar fixation pin 2pk 40mm, quick release.The product associated with the reported event is within the scope of recall z-0021-2022.However, there is insufficient information to evaluate whether the subject issue of the recall was a cause or contributor to the reported event.The device was not returned for evaluation and no medical or other records containing treatment information or patient information have been received; therefore, the reported event cannot be confirmed, nor can the circumstances or potential causes or contributors to the alleged event be evaluated.Should additional, material information become available that permits more analysis or conclusions, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
Legal case - usa (mdl no.3044 (ngg) (mmh).Related case - (b)(4) rk rev 1, (b)(4) rk rev 2, (b)(4) rk pain.It was reported that approximately 79 months after a left total knee replacement procedure, the patient has experienced prosthesis wear, pain, swelling, decreased range of motion, mental anguish, emotional distress, fear, loss of enjoyment of life, medical expenses, and financial losses.No further issues or complications were reported.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear, loss of range of motion or due to inclusion of the polyethylene in the packaging recall.However, the reported prosthesis wear could not be confirmed and potential contributions of user, or patient-related considerations to the event could not be assessed as the devices were not available for evaluation and no images, radiographs, or relevant clinical information was provided.
|
| |
|