| Model Number |
8888145039P |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, dialysis was in progress, the tape on the catheter was removed, and blood spurted out onto the pillow.The leak was at the extension tube at the red luer adapter.The dialysis was aborted.The renal reg was advised.The patient was to present the next day for catheter removal, new catheter insertion, and dialysis.There was no reported patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the catheter was flushed and a leak was present in the venous extension tube.The placement of the cut suggests it was created by clamping the tubing too close to the luer adapter.Clamping the extension tube close to the luer adapter can cause excess stress on the extension tubing which can lead to breaks/leaks in the tubing where it connects to the luer adapter.It was reported that there was a leak or hole on the extension tube at or near the luer adapter.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: clamping repeatedly on the same spot could weaken the tubing and clamping near the adapter and hub should be avoided.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|