| Catalog Number |
72201492 |
| Medical Device Problem Code |
Failure to Fire (2610)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, when the surgeon activated the second suture of the ultra fast-fix did not release and remained stuck.The suture was removed with tweezers.The procedure was completed using a back-up device.There was a delay less than 30 minutes and no further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference case (b)(4).A device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, when the surgeon activated the second suture of the ultra fast-fix did not release and remained stuck.The suture was removed with tweezers.The procedure was completed using a back-up device.There was a delay less than 30 minutes and no further complications were reported.Patient current status is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).B5: description updated.
|
| |
|