| Catalog Number |
72201772 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during surgery, the biosure-ha screw broke when turned into the tunnel.The broken pieces were removed from the patient.Surgery was completed with a back-up device instead.It is unknown if there was any surgical delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H10 h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for evaluation.
|
| |
|
Search Alerts/Recalls
|