MAUDE Adverse Event Report: TORNIER INC AEQUALIS FLEX REVIVE THREADED ROD; FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TORNIER INC AEQUALIS FLEX REVIVE THREADED ROD; FORCEPS Back to Search Results
Catalog Number ARS779600
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/12/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.
 
Event or Problem Description
It was reported that the threads on the adapter and rod cold welded together.It was also assumed that the threads were off a little and force from slap hammer caused this.
 
Additional Manufacturer Narrative
The reported event was not confirmed, since the device was not returned for evaluation and no other evidences were provided.A device inspection was not possible since the affected device was not returned, and no other evidences were provided for investigation.The device history record could not be reviewed because the affected device was not returned, and the lot number was not communicated.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.A review of the labeling did not indicate any abnormalities.More information as well as the affected device must be available in order to determine the exact root cause of the alleged failure.If any further information is provided, the complaint report will be updated.
 
Event or Problem Description
It was reported that the threads on the adapter and rod cold welded together.It was also assumed that the threads were off a little and force from slap hammer caused this.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AEQUALIS FLEX REVIVE THREADED ROD
Common Device Name
FORCEPS
Manufacturer (Section D)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer (Section G)
TORNIER INC
10801 nesbitt avenue s
bloomington MN 55437
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key19978738
Report Number0001649390-2024-00363
Device Sequence Number8234306
Product Code HTD
UDI-Device Identifier00846832069071
UDI-Public00846832069071
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue NumberARS779600
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/15/2024
Supplement Date Received by Manufacturer11/07/2024
Initial Report FDA Received Date08/13/2024
Supplement Report FDA Received Date12/05/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-