| Model Number |
CR2 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/26/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not power on.In this state, the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient use associated with the reported event.
|
| |
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device would not power on.In this state, the device would be inoperable and defibrillation therapy would not be available if needed.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker evaluated the customer's device and verified the reported issue.It was observed that the device was able to power on using the on/off button.Stryker determined that the cause of the reported issue was due to the lid switch, designator sw5.Stryker archived the device.The customer received a replacement device.
|
| |
|