• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RETROGRADE DRL 8.5MM; BIT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RETROGRADE DRL 8.5MM; BIT, SURGICAL Back to Search Results
Catalog Number 72204043
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an arthroscopy, the retrograde 8.5 drill bit broke when activating the helix inside the femoral tunnel since the doctor did not follow the instructions correctly.No part was left inside the patient.The procedure was completed with a surgical delay of less than 30 minutes using a smith and nephew back up device.No further complications were reported.
 
Additional Manufacturer Narrative
H10: internal complaint reference number: (b)(4).
 
Additional Manufacturer Narrative
H10: h2: additional information on: a1, a3, b5 & h6 (impact code).
 
Event or Problem Description
It was reported that during an arthroscopy, the retrograde 8.5 drill bit broke when activating the helix inside the femoral tunnel since the doctor did not follow the instructions correctly.No part was left inside the patient, all was removed with tweezers.The procedure was completed with a surgical delay of less than 30 minutes using a smith and nephew back up device in the initial bone hole.No further complications were reported.
 
Additional Manufacturer Narrative
A device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for precautions and warnings related to the use of the device.Based on this investigation, the need for corrective action is not indicated.If any additional information related to this event is received in the future, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETROGRADE DRL 8.5MM
Common Device Name
BIT, SURGICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key19981245
Report Number1219602-2024-01742
Device Sequence Number18163379
Product Code GFG
UDI-Device Identifier00885554037722
UDI-Public885554037722
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/17/2025
Device Catalogue Number72204043
Device Lot Number4997177
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2024
Supplement Date Received by Manufacturer08/20/2024
09/06/2024
Initial Report FDA Received Date08/14/2024
Supplement Report FDA Received Date08/24/2024
09/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/17/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-