| Model Number |
MC2-5590S |
| Medical Device Problem Codes |
Break (1069); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was returned to newcastle university; no product was returned to the manufacturer for investigation.A root cause was unable to be determined with the information provided.
|
| |
|
Event or Problem Description
|
|
An investigation report from an implant retrieval center was received and reported that during examination, the rod failed force testing.The cage and rolling elements, keeper plate, washers, extending bar, radial bearing and outer casing were covered in debris.The radial bearing was in place and it could rotate.The drive pin was also intact.It is unknown why the rod was removed.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the dhr documents indicates the device was manufactured by the specified requirement at the time and met all the required inspections upon release to finished goods.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|