|
The investigation has determined that lower than expected vitros intact pth ii (ipth ii) results were obtained from several different patient samples when tested using vitros ipth ii reagent on a vitros xt 7600 integrated system.The samples were considered discordant when compared to a non-vitros diasorin method.A definitive assignable cause of the lower than expected vitros ipth result could not be determined.Vitros ipth ii reagent was a new reagent for the customer which is currently being validated to bring testing in house.Therefore, no previous vitros ipth ii reagent lots of qc testing were available for evaluation.However, based on quality control (qc) results processed after the events, a vitros ipth ii lot 0040 and lot 0050 performance issue is not a likely contributor to the event.A diagnostic precision test using vitros total thyroid control (ttc) fluids was not performed to confirm the performance of the vitros analyzer.Therefore, an instrument cannot be entirely ruled out as a potential contributor to the events.Furthermore, it was not possible to establish if the customer was following the sample collection device manufacturer's recommendation for centrifugation protocol; therefore, pre-analytical sample processing cannot be ruled out as a contributing factor.It is possible that cellular debris due to poor sample preparation was present in the affected samples, although this cannot be confirmed.Additionally, the ortho tsc were able to determine that the patient's samples were stored frozen between the testing events and processed at the reference lab the same day or the following day.As per the vitros ipth ii ifu, specimen collection and preparation storage section: "intact pth is labile and susceptible to fragmentation.Correct handling of patient samples is necessary to ensure that the pth molecule remains intact.The degree of fragmentation will depend on both time and temperature of storage." however, no information was provided with regards to the handling of the patient's samples on the day of the event or at the reference lab, therefore, it is possible improper handling of the patient's samples contributed to the lower than expected vitros ipth ii lot 0040 and lot 0050 results, although this could not be confirmed.Continual tracking and trending of complaints has not identified any signals that would indicate a potential systematic issue with vitros ipth reagent lot 0040 and lot 0050.
|